My current system requires periodic optical calibration to ensure performance; how does the xxpress® system address this?

/My current system requires periodic optical calibration to ensure performance; how does the xxpress® system address this?

My current system requires periodic optical calibration to ensure performance; how does the xxpress® system address this?

2018-01-22T15:48:48+00:00 December 4th, 2017|

The xxpress® optical system components do not move during operation, with the exception of the filter wheel, which moves while measuring the five emission wavelengths for each sample at each cycle.

The optical system is calibrated prior to shipment and we recommend that a recalibration is performed after 2500 runs or 1 year whichever occurs sooner. Pricing for this calibration plan is available from your local representative.